Z-0796-2021 Class II Terminated

Recalled by Steris Corporation — Mentor, OH

FDA device recall Z-0796-2021 was initiated by Steris Corporation on Dec 2, 2020 and is designated Class II. Reason for recall: Purple ink located on the outside label of Vis-U-All Pouches may migrate ( bleed ) following steam processing causing cosmetic defects on the exterior of the pouch and in some instances, transference… The recall status is terminated (terminated Jun 21, 2022). Affected quantity: 229 cases.

Recall details

Product Type
Devices
Report Date
Jan 13, 2021
Initiation Date
Dec 2, 2020
Termination Date
Jun 21, 2022
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
229 cases

Product description

Vis-U-All High Temp 5.5"x10" Self Seal Pouch 200 pouches per box; 4 boxes per case Model Number: 884510

Reason for recall

Purple ink located on the outside label of Vis-U-All Pouches may migrate ( bleed ) following steam processing causing cosmetic defects on the exterior of the pouch and in some instances, transference of the ink onto the device(s) contained within the pouch

Distribution pattern

Nationwide Foreign:

Code information

Lot Number: 200435A, 200635A

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0796-2021
openFDA file date
Nov 24, 2026
Loaded by RegDataLab
Nov 25, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Steris Corporation? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works