Z-0836-2015 Class II Terminated
FDA device recall Z-0836-2015 was initiated by Ion Beam Applications S.A. on Nov 7, 2014 and is designated Class II. Reason for recall: Two issues were identified: 1. In the release of adaPT insight v1.3.2., the kV/kV Single Source acquisition workflow was linked to a wrong geometrical calibration file, resulting in an… The recall status is terminated (terminated Jun 29, 2015). Affected quantity: 1.
Recall details
- Product Type
- Devices
- Report Date
- Dec 31, 2014
- Initiation Date
- Nov 7, 2014
- Termination Date
- Jun 29, 2015
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 1
Product description
adaPT insight (12C), v1.3.2, PAT.109 (US), Treatment room 4: I2C is used with a charged particle or photon radiation therapy system for localization of the patient position with respect to the therapy equipment and to provide correction feedback to the radiation therapy device.
Reason for recall
Two issues were identified: 1. In the release of adaPT insight v1.3.2., the kV/kV Single Source acquisition workflow was linked to a wrong geometrical calibration file, resulting in an alignment offset for this specific mode of acquisition. 2. A gantry collision incident required the IBA service team to verify the gantry alignment and protons I x-ray beams colinearity, du
Distribution pattern
US Distribution in PA only.
Code information
PAT.109 (US), Treatment room 4