Z-1492-2014 Class II Terminated
FDA device recall Z-1492-2014 was initiated by Ion Beam Applications S.A. on Dec 4, 2013 and is designated Class II. Reason for recall: Internal testing of the Proteus 235 Proton Therapy System revealed incorrect dose rate calculation in case of beam line option other than Zero (0). The recall status is terminated (terminated Jun 12, 2014). Affected quantity: 4.
Recall details
- Product Type
- Devices
- Report Date
- Apr 30, 2014
- Initiation Date
- Dec 4, 2013
- Termination Date
- Jun 12, 2014
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 4
Product description
Proteus 235 Proton Therapy System Legacy Software version 7.2.7 and higher, models 110, 112, 113, and 116. Proton therapy
Reason for recall
Internal testing of the Proteus 235 Proton Therapy System revealed incorrect dose rate calculation in case of beam line option other than Zero (0).
Distribution pattern
US Distribution including the states of OK, IL, NJ, and WA.
Code information
Proteus 235 Proton Therapy System Legacy Software version 7.2.7 and higher, models 110, 112, 113, and 116.