Z-2097-2017 Class II Terminated

Recalled by Ion Beam Applications S.A. — Louvain La Neuve

FDA device recall Z-2097-2017 was initiated by Ion Beam Applications S.A. on Apr 10, 2017 and is designated Class II. Reason for recall: A PBS (Pencil Beam Scanning) irradiation may pause for different reasons. After a pause, irradiation may restart from the beginning instead of recalculating the field based on the already delivered… The recall status is terminated (terminated Feb 16, 2018). Affected quantity: 2.

Recall details

Product Type
Devices
Report Date
May 24, 2017
Initiation Date
Apr 10, 2017
Termination Date
Feb 16, 2018
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
2

Product description

Proton therapy system -Proteus 235 aka Proteus Plus. A medical device designed to produce and deliver a proton beam for the treatment of patients with localized tumors and other conditions susceptible to treatment by radiation.

Reason for recall

A PBS (Pencil Beam Scanning) irradiation may pause for different reasons. After a pause, irradiation may restart from the beginning instead of recalculating the field based on the already delivered dose.

Distribution pattern

FL

Code information

Serial numbers: PAT.003, PAT.006

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2097-2017
openFDA file date
Nov 16, 2026
Loaded by RegDataLab