Z-2178-2025 Class II Ongoing

Recalled by Steris Corporation — Mentor, OH

FDA device recall Z-2178-2025 was initiated by Steris Corporation on Jun 23, 2025 and is designated Class II. Reason for recall: The DFPM yoke assembly ("yoke assembly") may detach from the spring arm. Should the yoke assembly detach from the spring arm, the yoke assembly would not be expected to fully detach but rather hang… The recall status is ongoing. Affected quantity: 59 units.

Recall details

Product Type
Devices
Report Date
Aug 6, 2025
Initiation Date
Jun 23, 2025
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
59 units

Product description

Dual Flat Panel Monitor (DFPM) accessories for HarmonyAIR E-Series Surgical Lighting Systems

Reason for recall

The DFPM yoke assembly ("yoke assembly") may detach from the spring arm. Should the yoke assembly detach from the spring arm, the yoke assembly would not be expected to fully detach but rather hang from the wiring within the yoke assembly.

Distribution pattern

Worldwide - US Nationwide and Puerto Rico and the country of Canada.

Code information

UDI: 00724995171155: Serial numbers: 0416920048, 0417820084, 0417820088, 0417820096, 0417820110, 0417820092, 0417820100, 0418420047, 0419620004, 0419520013, 0428620109, 0428620112, 0430020191, 0431720044, 0435120043, 0435120047, 0435120051, 0401321042, 0413221047, 0413221199, 0413721008, 0413721012, 0417221023, 0417221027, 0417221031, 0420021018, 0420021022, 0420021026, 0420721021, 0429720003, 0420921073, 0426321186, 0426321190, 0427121121, 0427121125, 0427321012, 0427321002, 0427321005, 0413921017, 0434121095, 0434121098, 0434121101, 0434121104, 0434121107, 0434121110, 0434121113, 0400722081, 0406022021, 0406022025, 0406022034, 0406022038, 0406022017, 0406022014, 0406022011, 0417122078, 0417122081, 0417122084, 0417122087, 0417122090.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2178-2025
openFDA file date
Nov 16, 2026
Loaded by RegDataLab
Nov 17, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-2178-2025. Report a data issue: hello@regdatalab.com.