Z-2248-2016 Class II Terminated

Recalled by Steris Corporation — Mentor, OH

FDA device recall Z-2248-2016 was initiated by Steris Corporation on May 23, 2016 and is designated Class II. Reason for recall: STERIS has identified that the pivot pin assembly located on the table top may become loose preventing the table from articulating properly. The recall status is terminated (terminated Aug 22, 2016). Affected quantity: 148.

Recall details

Product Type
Devices
Report Date
Jul 27, 2016
Initiation Date
May 23, 2016
Termination Date
Aug 22, 2016
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
148

Product description

OT1000 Series Orthopedic Surgical Tables

Reason for recall

STERIS has identified that the pivot pin assembly located on the table top may become loose preventing the table from articulating properly.

Distribution pattern

AL;AK, AZ; AR; CA; CO; CT; DC; FL; GA; IL; IN; IA; NY; NJ; NC; OH; OK; OR; PA; PR; SC; TX; UT and VA. Foreign accounts: Canada; France; Korea; Panama; Qatar and Thailand.

Code information

OT1000, OT1100, and OT1200 Orthopedic Surgical Table

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2248-2016
openFDA file date
Nov 24, 2026
Loaded by RegDataLab
Nov 25, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Steris Corporation? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works