Z-2325-2012 Class III Terminated
FDA device recall Z-2325-2012 was initiated by Steris Corporation on Aug 2, 2012 and is designated Class III. Reason for recall: On 7/20/2012 the firm became aware that a lot of tubing was distributed by the supplier after it was rejected by Steris due to failing process indicators. The recall status is terminated (terminated Sep 4, 2013). Affected quantity: 77 packs.
Recall details
- Product Type
- Devices
- Report Date
- Sep 12, 2012
- Initiation Date
- Aug 2, 2012
- Termination Date
- Sep 4, 2013
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 77 packs
Product description
100 ft roll of tubing inside a cardboard box. Medical devices are sealed into the tubes prior to ethylene oxide sterilization. Tube maintains sterility of the enclosed medical devices until opened.
Reason for recall
On 7/20/2012 the firm became aware that a lot of tubing was distributed by the supplier after it was rejected by Steris due to failing process indicators.
Distribution pattern
Worldwide Distribution--USA (nationwide) including the states of AL, AZ, CA, FL, IL, IN, KY, LA, MO, MS, NJ, OH, PA, SD, TX, and VA., and the countries of Japan, Peru, and Suriname.
Code information
Product code KCT; Lot number: 60263
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-2325-2012