Z-2536-2024 Class II Ongoing

Recalled by Steris Corporation — Mentor, OH

Recall Details

Product Type
Devices
Report Date
August 14, 2024
Initiation Date
June 13, 2024
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
37

Product Description

Hookup Model/Catalog Number: 2-8-540 Product Description: Hookups provide the interface between an Advantage Plus" Automated Endoscope Reprocessor and Advantage Plus" Pass-Thru Automated Endoscope Reprocessor ( AER ) and endoscopes, which allow perfusion of liquids dispensed by the AER to contact and high-level disinfect the channels of flexible endoscopes.

Reason for Recall

IFU for hookup were incorrectly reflected within hookup IFUs and online reference system, Hookup Lookup (HULU). Use of an incorrect hookup connection and/or parameter set may prevent an endoscope from being properly disinfected, which may present a risk of patient contamination or infection

Distribution Pattern

Worldwide distribution - United States Nationwide and the countries of Canada, Belgium, Czech Republic, Denmark, Germany, Spain, Italy, Slovenia, Slovakia, Australia, Israel, Jordan, Kuwait, Namibia, Saudi Arabia, South Africa, Cote d Ivoire, Afghanistan, Albania, Belarus, Botswana, Kazakhstan, Malaysia, Qatar, Ukraine, United Arab Emirates, Nicaragua, and Nigeria.

Code Information

Lot Code: UDI-DI 00677964091223; Lot Numbers 404360, 415307, 421967, 423540, 425801, 427239, 435309, 436068, 447102, 516866, 523989, 539379, 540599, 542619, 544723, 550675, 554204, 560426, 587537, 588583, 594221, 604965, 61450